Do AO2020 and AO2040 measure different gases?
The designation mainly distinguishes the mechanical construction. The installed modules and their configuration determine the measurement.
A modular platform that builds the measurement around the process composition.

Advance Optima AO2000 combines analytical modules with shared management. The AO2020 19-inch rack and AO2040 wall-mounted versions identify the system construction, not the gas being measured.
| Module | Principle | Example application |
|---|---|---|
| Uras26 | Non-dispersive infrared · NDIR | CO, CO₂ and other IR-active components |
| Limas21 UV / HW | UV photometry / hot-wet version | Selected components for process and flue-gas treatment |
| Magnos28 | Paramagnetic | Oxygen |
| Caldos25 / Caldos27 | Thermal conductivity | Composition of defined mixtures |
| Fidas24 | Flame ionisation · FID | Hydrocarbons |
| LS25 | TDLS laser | In-situ measurement in compatible configurations |
Specify the module, measured components, ranges, wetted materials, calibration gases and interfaces separately. For replacement, AO2020 or AO2040 alone is insufficient: provide the serial number, module configuration and sample data.
For flammable mixtures, check the specific system version and permitted gas conditions as well as the installation area classification.
To select AO2000, send us the following information, even if preliminary:
The designation mainly distinguishes the mechanical construction. The installed modules and their configuration determine the measurement.
The response depends on the complete mixture. It suits defined mixtures; changes in the background gas must be assessed.