Measurement quality starts at the sampling point. An accurate analyser cannot correct a sample altered during transport.

ABB ESTRATTIVO

Keep the sample representative

The system delivers a sample compatible with the analyser cell while preserving the components of interest. Select the approach before sizing the cabinet, utilities and drains.

FunctionSelection criterion
Sampling and filtrationRepresentativeness, dust loading and maintenance access.
Sample lineMaterials, length, internal volume and heating requirements.
Pressure reductionPermitted pressure, stability and condensation risk.
Moisture treatmentCompatibility of cooling with the components to be measured.
Flow regulationRequired flow, monitoring and low-flow alarm.
CalibrationTest gas injection point and the section of the measurement chain being checked.
Venting or recoveryBackpressure, gas destination and condensate collection.

Cold-dry or hot-wet?

Cold-dry

The sample is cooled and condensate separated. Check that treatment does not remove components of interest and that the measurement basis matches the required data.

Hot-wet

The sample path is kept hot to prevent condensation. Temperature must be controlled along the entire line, including fittings and the analyser inlet.

Do not add a cooler by default: soluble or readily adsorbed components require a dedicated assessment.

Response time and leak integrity

An initial estimate of transport delay is the line volume divided by volumetric flow at actual conditions. Add the instrument response and the volumes of filters, vessels and switching assemblies.

For low-concentration oxygen, even slight air ingress can significantly affect the reading. Check leak integrity, fittings, materials and pressure along the entire path.

Geyser IC integration

We work with you to define the sampling point, sample schematic and scope of supply for panels and cabinets. The scope can include P&IDs, bills of materials, connections and agreed functional checks.

Explore panel and SCS integration

Information for a quotation

To select sample conditioning, send us the following information, even if preliminary:

  • Components to be measured, background gas and expected interferents.
  • Normal, minimum and maximum ranges; units and dry or wet basis.
  • Pressure and temperature at the sampling point; moisture, dust and condensable components.
  • Required response time, number of points and distance from the sampling point.
  • Area classification, indoor/outdoor installation, available signals and utilities.
  • Available P&ID, layout, gas and condensate destinations, utilities and scope boundaries.

Frequently asked questions

Is a cooler always necessary?

No. It depends on the technology and components. Some applications require the sample to remain hot.

Does instrument calibration also check the sample line?

That depends on where the test gas is introduced. A test at the analyser inlet does not automatically check the entire sampling path.